Measurement |
4 Terminal Kelvin/Thomson principle eliminates error due to lead resistance. Open circuit measurement voltage~2V dc |
|
Display |
0.8" LED 6,000 count with automatic decimal point and polarity indication |
|
Ranges |
6 push button selected with LED indication |
|
Calibration |
Digital, security key protected |
|
Protection |
415rms maximum at input terminals |
|
Battery |
Sealed lead acid, rechargeable cells giving a minimum of 1 hour of continuous measurement on the lowest 10 amp ranges and 20 hours on all other ranges. Internal charger with battery state indicator |
|
Terminals |
6mm binding posts accept spade tags and 4mm banana plugs |
|
Working Temp |
0...40°C rel. humidity 80% max. non-condensing |
|
Storage Temp |
-20...+50°C |
|
Mains Supply |
100/120/220/240 Volts +10% -13% 47Þ63Hz 20VA |
|
Safety |
IEC 1010 Protective Class 1 |
|
EMC |
Complies with EN61326 |
|
Dimensions |
343 x 327 x 152mm (W D H) approx |
|
Mass |
8 kg approx |
|
Accessories |
Supplied complete with mains cord, measuring leads with Kelvin Clips, handbook and calibration certificate |
|
Range |
Resolution |
Typical current |
Accuracy @ 20°C ±5 °C, 1 year |
Temperature Coefficient / °C |
|
60Ohm |
10mOhm |
1mA |
±(0.15% Rdg +0.05% FS) |
40 ppm Rdg + 30 ppm FS |
|
6Ohm |
1mOhm |
10mA |
|
|